Introduction
Calculate the combined standard uncertainty using the JCGM 100:2008 (GUM) method. Combining uncertainties using the GUM method is a core skill for every calibration and testing laboratory. Most labs know the formula but struggle when correlated inputs are involved. When you get it wrong, the error affects the expanded uncertainties reported on ISO/IEC 17025 accredited certificates and, where applicable, scopes of accreditation.
Mistakes rarely come from the formula itself. They come from the inputs, including skipped divisors, missing sensitivity coefficients, and treating correlated contributors as independent.
This guide shows you how to calculate combined standard uncertainty per the GUM for both uncorrelated and correlated input quantities. You’ll find step-by-step procedures, three worked examples from real calibrations, Excel formulas you can copy, and a list of common mistakes to check your work against. Make sure to check out the end of the guide for answers to frequently asked questions.
Key Takeaways
- Calculate combined standard uncertainty with the root sum of squares, which the GUM calls the law of propagation of uncertainty (JCGM 100:2008, § 5.1.2).
- Convert every contributor to a standard uncertainty in the unit of the result first, using its divisor and sensitivity coefficient.
- Add covariance terms when input quantities are correlated (JCGM 100:2008, § 5.2.2).
- Treat the combined standard uncertainty as one standard deviation, about 68 % for a normal distribution. Multiply it by a coverage factor (k) to get the expanded uncertainty.
- Use =SQRT(SUMSQ(…)) in Excel when all sensitivity coefficients equal one.
Table of Contents
Click the links below to skip to any section.
What is Combined Standard Measurement Uncertainty
According to the JCGM 200:2012 – Vocabulary in Metrology (VIM), combined standard measurement uncertainty is the “standard measurement uncertainty that is obtained using the individual standard measurement uncertainties associated with the input quantities in a measurement model.”
In simple terms, it is a standard measurement uncertainty calculated by combining the variances of individual standard uncertainties from the input quantities in a measurement model.
In the image below is the definition from the JCGM 200, § 2.31.
Why Combining Uncertainties Uses the RSS Method
The RSS method is the root sum of squares method, which is calculated as the square root of the sum of squares. It is also called adding in quadrature, or “summation in quadrature” in NPL’s A Beginner’s Guide to Uncertainty of Measurement (Bell, 1999).
Uncertainties are combined using this method because standard uncertainties (i.e. standard deviation equivalents) must be converted to variances before they can be added together. You cannot combine standard deviations using linear addition. Hence, the sum of squares part of the RSS method.
Like the Pythagorean theorem and vector addition, independent contributors combine like the perpendicular sides of a right triangle: the combined uncertainty is the total magnitude or length of the triangle’s hypotenuse.
Also, each contributor has its own probability distribution. When the distributions are combined, the resulting distribution tends to be approximately normal according to the Central Limit Theorem, as long as certain conditions are met. See JCGM 100:2008, § G.2 and § G.6.6. Otherwise, the resulting distribution could be non-normal. For example, when a rectangular distribution dominates the uncertainty analysis, the resulting distribution approaches rectangular and the coverage factor for approximately 95 % level of confidence is 1.65.
How to Calculate Combined Standard Uncertainty
Choose one of the two methods from JCGM 100:2008, depending on whether the input quantities are independent (i.e., uncorrelated) or interdependent (i.e., correlated).
| Input Quantities | Method |
|---|---|
| Uncorrelated (r = 0) | Uncorrelated Input Quantities |
| Correlated (r ≠ 0) | Correlated Input Quantities |
In simple terms, input quantities are uncorrelated if their covariance and correlation coefficient are zero. Independent inputs are always uncorrelated, which is the most common scenario.
- Independence – “Two random variables are statistically independent if their joint probability distribution is the product of their individual probability distributions” (JCGM 100:2008, § C.3.7).
- Correlation – “The relationship between two or several random variables within a distribution of two or more random variables” (JCGM 100:2008, § C.2.8).
A non-zero correlation coefficient doesn’t necessarily mean the correlation is significant, and JCGM 100:2008 does not specify an r threshold for significance.
When in doubt, estimate r and calculate the combined standard uncertainty uc with and without the covariance terms. If the difference is negligible (e.g., it doesn’t change your reported uncertainty after rounding), treat the inputs as uncorrelated.
Uncorrelated Input Quantities
According to JCGM 100:2008, § 5.1.2, the combined standard uncertainty is the positive square root of the combined variance.
This is where the Root Sum of Squares (RSS) acronym comes from. The GUM calls this formula the law of propagation of uncertainty.
IMPORTANT: Use this method only when the input quantities are independent. Otherwise, use the correlated method (see next section). Inputs are typically correlated when they share a common source, such as the same reference standard, instrument, or environmental influence.
Quick example: Three uncorrelated contributors of 5 µV/V, 2 µV/V, and 3 µV/V, each with ci = 1, combine to √(5² + 2² + 3²) = √38 = 6.2 µV/V. Adding them linearly gives 10 µV/V and overstates the uncertainty by 62 %.
Calculate the combined standard uncertainty for uncorrelated input quantities per JCGM 100:2008, § 5.1.2, following the instructions below.
- Determine all the input terms:
- Determine the measurement uncertainty of each contributor (i.e. input quantity).
- Convert each uncertainty to a standard uncertainty (i.e. standard deviation equivalent based on its probability distribution).
- Determine the sensitivity coefficient for every uncertainty contributor, evaluated at the value of each input xi, keeping its sign (+, −).
- Calculate the Combined Variance:
- Calculate the uncertainty component ui(y) by multiplying each standard uncertainty u(xi) by its sensitivity coefficient ci.
- Convert each uncertainty component ui(y) to a variance by squaring it.
- Calculate the combined variance uc2(y) by adding together all the variances.
- Calculate the Combined Standard Uncertainty uc(y) by taking the square root of the combined variance.
Note 1: Sensitivity coefficients typically equal one (ci = 1) when a contributor is expressed in the same unit as the measurand, but it depends on the model function. For example, averaging two readings gives ci = ½. When in doubt, calculate them from the model (the partial derivative above, or numerically by changing xi by a small amount and observing the change in y), or determine them experimentally when no model is available.
Note 2: This method uses a first-order (linear) approximation of the model at each measurement point. It is suitable for most calibrations. If the model is significantly nonlinear relative to the size of the uncertainties, higher-order terms may be needed (JCGM 100:2008, § 5.1.2 Note), and a Monte Carlo method (JCGM 101:2008) is often a better tool.
Correlated Input Quantities
According to JCGM 100:2008, § 5.2.2, when input quantities are correlated, the combined variance includes a covariance term for each correlated pair. Depending on the signs (+,-), correlation can increase or decrease the combined standard uncertainty.
IMPORTANT: Use this method only when the input quantities are correlated. Otherwise, use the uncorrelated method (see previous section). Inputs are typically correlated when they share a common source, such as the same reference standard, instrument, or environmental influence.
Calculate the combined standard uncertainty for correlated input quantities per JCGM 100:2008, § 5.2.2, following the instructions below.
- Determine all the input terms:
- Determine the measurement uncertainty of each contributor (i.e. input quantity).
- Convert each uncertainty to a standard uncertainty (i.e. standard deviation equivalent based on its probability distribution).
- Determine the sensitivity coefficient for every uncertainty contributor, evaluated at the value of each input xi, keeping its sign (+, -).
- Determine the correlation coefficient r(xi,xj) for each pair of correlated contributors, using the equation above or engineering judgment.
- Calculate the Total Variance term:
- Calculate the uncertainty component ui(y) by multiplying each standard uncertainty u(xi) by its sensitivity coefficient ci.
- Convert each uncertainty component ui(y) to a variance by squaring it.
- Calculate the total variance by adding together all the variances.
- Calculate the Total Covariance term:
- For each pair of correlated contributors (r ≠ 0), calculate the covariance term by multiplying the following:
- the two sensitivity coefficients (ci,cj),
- the two standard uncertainties (u(xi),u(xj)), and
- their correlation coefficient (r(xi,xj)).
- Calculate the total covariance by adding together all covariance terms, keeping the signs (+, -) and counting each pair only once (i < j).
- For each pair of correlated contributors (r ≠ 0), calculate the covariance term by multiplying the following:
- Calculate the Combined Standard Uncertainty:
- Multiply the total covariance term by 2.
- Add together the total variance and twice the total covariance terms.
- Calculate the square root of the result to get uc(y).
Note 1: If every correlated pair is fully correlated (r = +1), you can add the ci • u(xi) values directly and take the absolute value. This gives the same uc as steps 2 to 4.
Note 2: Note 2 in the previous section (first-order approximation) also applies here.
Examples of Calculating Combined Standard Uncertainty
Review the examples below. They include scenarios for calculating the combined standard uncertainty for both uncorrelated and correlated uncertainty contributors. For a complete budget from start to finish, see this uncertainty budget example.
Uncorrelated Example: DC Current Measured with a Shunt and a DMM
In this example, DC current is calculated via Ohm’s law from the voltage drop measured across a calibrated resistor (i.e., shunt).
| Formula | Input Terms |
|---|---|
|
I = \frac{E_{M}}{R_{S}}
|
I = Calculated DC Current
EM = Measured DC Voltage from Digital Multimeter RS = Certified Resistance of Reference Shunt |
The table below gives the input quantities for this example.
| Quantity | Value |
|---|---|
| Nominal Source Current (Inom) | 10 A |
| Shunt Resistance (RS) | 0.01 Ω (10 mΩ) |
| DMM Voltage Reading (EM) | 100 mV |
| Result: Measured Current (I) | 100 mV / 0.01 Ω = 10 A |
The digital multimeter (DMM) and the shunt are separate instruments, each calibrated independently. They do not share any error sources, so all input quantities are uncorrelated.
Step 1: Determine the input terms
This evaluation includes uncertainty contributors for two separate instruments, each with its own sources of uncertainty. Additionally, the uncertainties will have different units of measurement. Therefore, one of the following approaches must be used to avoid errors in the final estimated uncertainty:
- Propagate each uncertainty through the measurement function (one factor at a time). Change one input by its standard uncertainty, recalculate the result, and record the change in amperes. No sensitivity coefficients are needed.
- Use sensitivity coefficients. Determine each coefficient from the measurement function, then multiply each standard uncertainty by its coefficient to convert it to amperes.
- Use relative uncertainties. Express each uncertainty as a percentage of its input. This works when the measurement function is a product or quotient, like I = EM/RS.
This example uses sensitivity coefficients (i.e., Option b) to convert input quantities to the same unit of measure before calculating the combined standard uncertainty.
In the table below, sensitivity coefficients are determined from the measurement model for the voltage and resistance terms.
| Coefficients | Formula | Sensitivity Coefficient |
|---|---|---|
| Voltage Coefficient (cE) |
\frac{\partial I}{\partial E_{M}} = \frac{1}{R_{S}} = 100\ \text{A/V}
|
0.1 mA/µV |
| Resistance Coefficient (cR) |
\frac{\partial I}{\partial R_{S}} = \frac{- E_{M}}{R_{S}^{2}} = - 1000\ \text{A}/\Omega
|
-1 mA/µΩ |
Next, the uncertainty contributors associated with the measurement model are identified, quantified, characterized, and converted to standard uncertainties (i.e., standard deviation equivalents).
The table below includes a summary of each uncertainty contributor along with its uncertainty value, probability distribution, divisor (based on the probability distribution), standard uncertainty, and sensitivity coefficient.
| Contributor | Value | Distribution | Divisor | u(xi) | ci |
|---|---|---|---|---|---|
|
Repeatability
(s = 0.30 mA, n = 10) |
0.30 mA | Normal | √10 | 0.095 mA | 1 mA/mA |
| Reproducibility | 0.55 mA | Normal | 1 | 0.550 mA | 1 mA/mA |
| DMM Calibration Uncertainty
(U = 0.16 µV, k = 2) |
0.16 µV | Normal | 2 | 0.080 µV | 0.1 mA/µV |
|
DMM Drift
±(9 µV/V rdg + 3 µV/V range) |
1.2 µV | Rectangular | √3 | 0.693 µV | 0.1 mA/µV |
|
DMM Resolution
(0.01 µV) |
0.01 µV | Rectangular | 2√3 | 0.0029 µV | 0.1 mA/µV |
| DMM Temperature
(Zero within 18 °C to 28 °C) |
0.00 µV | Rectangular | √3 | 0.000 µV | 0.1 mA/µV |
| DMM Loading Effect
(EL = EDMM x RS / (RS + RDMM)) |
0.0001 µV | Rectangular | √3 | 0.0000577 µV | 0.1 mA/µV |
| Thermal EMF (connections)
(0.3 µV/°C × ±0.5 °C) |
0.15 µV | Rectangular | √3 | 0.087 µV | 0.1 mA/µV |
| Shunt Calibration Uncertainty
(U = 100 µΩ/Ω, k = 2) |
1.00 µΩ | Normal | 2 | 0.500 µΩ | -1 mA/µΩ |
| Shunt Drift
(±50 µΩ/Ω per year) |
0.50 µΩ | Rectangular | √3 | 0.289 µΩ | -1 mA/µΩ |
| Shunt Temperature Coefficient
(20 µΩ/Ω/°C × ±5 °C) |
1.00 µΩ | Rectangular | √3 | 0.577 µΩ | -1 mA/µΩ |
| Shunt Power Coefficient
(0.1 µΩ/Ω/W × ±1 W) |
0.001 µΩ | Rectangular | √3 | 0.000577 µΩ | -1 mA/µΩ |
Note: Repeatability and reproducibility are evaluated directly in amperes, so ci = 1. The other contributors are in volts or ohms and need the model to convert them to amperes.
Step 2: Calculate the combined variance
This step focuses on calculating the combined variance. To accomplish this, each input term’s standard uncertainty was multiplied by its respective sensitivity coefficient ci to calculate the uncertainty component ui(y). Next, each uncertainty component was squared to convert it to a variance ui2(y). Then, the variances were added together to calculate the combined variance.
A summary is provided in the table below along with each input term’s percentage of the total combined variance. The “% of Total” column shows which contributors are the most significant.
| Contributor | ci • u(xi) | [ci • u(xi)]2 | % of Total |
|---|---|---|---|
| Repeatability | 0.0949 mA | 9.000E-03 mA2 | 0.9 % |
| Reproducibility | 0.5500 mA | 3.025E-01 mA2 | 30.8 % |
| DMM Calibration Uncertainty | 0.0080 mA | 6.400E-05 mA2 | 0.0 % |
| DMM Drift | 0.0693 mA | 4.800E-03 mA2 | 0.5 % |
| DMM Resolution | 0.0003 mA | 8.333E-08 mA2 | 0.0 % |
| DMM Temperature | 0.0000 mA | 0.000E+00 mA2 | 0.0 % |
| DMM Loading Effect | 0.00000577 mA | 3.333E-11 mA2 | 0.0 % |
| Thermal EMF (connections) | 0.0087 mA | 7.500E-05 mA2 | 0.0 % |
| Shunt Calibration Uncertainty | -0.5000 mA | 2.500E-01 mA2 | 25.4 % |
| Shunt Drift | -0.2887 mA | 8.333E-02 mA2 | 8.5 % |
| Shunt Temperature Coefficient | -0.5774 mA | 3.333E-01 mA2 | 33.9 % |
| Shunt Power Coefficient | -0.000577 mA | 3.333E-07 mA2 | 0.0 % |
| Combined Variance | 0.9831 mA2 | 100.0 % | |
Note: The negative signs on the shunt contributors turn positive when squared. With uncorrelated inputs, the sign doesn’t affect the result.
Step 3: Calculate the combined standard uncertainty
Finally, calculate the combined standard uncertainty by computing the square root of the previously calculated combined variance.
A summary of the calculation is provided below.
Express the combined standard uncertainty in one of the following formats:
- absolute uncertainty – same unit as the result, or
- relative uncertainty – terms relative to the result.
Relative uncertainties (% or µA/A) make it easier to compare uncertainties across ranges and to carry them into the next step.
The table below shows the combined standard uncertainty in both absolute and relative forms.
| Result | Formula | uc(y) |
|---|---|---|
| Combined Standard Uncertainty (Absolute) |
u_{c}(I) = \sqrt{0.9831\ \text{mA}^{2}}
|
0.99 mA |
| Combined Standard Uncertainty (Relative) |
u_{c,rel}(I) = 100 \bullet 0.992\ \text{mA}\ /\ 10\ 000\ \text{mA}
|
0.0099 % |
| Combined Standard Uncertainty (Relative) |
u_{c,rel}(I) = 10^{6} \bullet 0.992\ \text{mA}\ /\ 10\ 000\ \text{mA}
|
99 µA/A |
Note: Express parts-per-million values with units, such as µA/A, rather than “ppm.” Notations like ppm, ppb and ppt are not SI units and are discouraged because they are ambiguous and language-dependent (SI Brochure, 9th ed., § 5.4.7; NIST SP 811, § 7.10.3).
Review the results
The table below ranks the most significant contributors to measurement uncertainty from most to least significant. Added together, the six contributors make up 99.99 % of the combined variance.
People often ask me, “How do I lower my measurement uncertainty?” The answer: find your largest contributors and reduce them first. That is where you’ll see the biggest improvement.
The results show that the shunt (i.e., resistor) is the largest contributor to measurement uncertainty. To reduce the uncertainty, consider one or more of the following:
- Controlling laboratory temperature to tighter limits
- Improving consistency in reproduced measurement results
- Using a shunt with a lower temperature coefficient, and/or
- Using a calibration supplier that can provide a smaller calibration uncertainty.
| Contributor (Most to Least Significant) | Share of combined variance |
|---|---|
| Shunt Temperature Coefficient | 33.9 % |
| Reproducibility | 30.8 % |
| Shunt Calibration Uncertainty | 25.4 % |
| Shunt Drift | 8.5 % |
| Repeatability | 0.9 % |
| DMM Drift | 0.5 % |
The table below shows the uncertainties grouped by source. The shunt and the Type A uncertainties dominate the uncertainty analysis. As a result, the contributions from the DMM are negligible.
| Contributors Grouped By Source (Most to Least Significant) | Share of combined variance |
|---|---|
| Shunt (calibration + drift + temperature coefficient) | 68 % |
| Repeatability + Reproducibility | 32 % |
| DMM (all relevant contributors combined) | <1 % |
Takeaway: Every contributor adds to the combined variance because the voltage and resistance come from independent instruments. With no shared error source, there is no covariance, and nothing cancels.
Correlated Example 2 (ISO 16063-21) below uses the same kind of ratio model, but the two voltages share an error source. See how the covariance term removes the shared reference standard error, while the independent drift remains.
Correlated Example 1: Absolute Pressure from Two Averaged Transducers
The Fluke molbox & molbloc-L Laminar Flow Element measurement function is a good example of correlated input quantities.
According to Fluke Technical Note 2011TN06B, the first evaluated uncertainty contributor is absolute pressure (i.e., L1). The formula in the document shows absolute pressure is the average of the upstream and downstream absolute pressure transducer measurements.
Fluke’s technical note presents absolute pressure as one combined uncertainty. This example breaks it into its individual contributors to show which ones are correlated. The uncertainty values are from the molbox1 A700K specifications and calibration report, at 500 kPa.
Measurement Function and Input Quantities
Below is the formula for calculating L1 – Absolute Pressure and a table with associated input quantities for each variable in a high-pressure system with A700K transducers.
| Formula | Input Terms |
|---|---|
|
\overline{P} = \frac{\left( P_{1} + P_{2} \right)}{2}
|
P1 = Upstream Absolute Pressure
P2 = Downstream Absolute Pressure \overline{P}= Mean Absolute Pressure |
The table below shows the values of the input terms. These are consistent with a high-pressure, upstream laminar flow element (i.e., before the unit under test) configuration.
| Parameter | Symbol | Value |
|---|---|---|
| Nominal Differential Pressure |
\Delta P
|
50 kPa |
| Upstream Absolute Pressure |
P_{1}
|
525 kPa |
| Downstream Absolute Pressure |
P_{2}
|
475 kPa |
| Mean Absolute Pressure |
\overline{P}
|
500 kPa |
Which Contributors Are Correlated?
Both transducers were calibrated against the same reference standard and are AutoZeroed against the same barometer, so those contributors are correlated (r = +1). Each transducer has its own drift and resolution, so those contributors are independent (r = 0).
| Contributor | Correlation |
|---|---|
| Reference Standard Uncertainty | Shared, r = +1 |
| AutoZero Reference | Shared, r = +1 |
| Drift | Independent, r = 0 |
| Resolution | Independent, r = 0 |
Sensitivity Coefficients
Based on the model formula (P̄ = (P1 + P2)/2), the sensitivity coefficients are c1 = c2 = ½.
In the table below, sensitivity coefficients are determined from the measurement model for the upstream and downstream pressure terms.
| Coefficients | Formula | Sensitivity Coefficient |
|---|---|---|
| Upstream Pressure Coefficient (cP1) |
\frac{\partial\overline{P}}{\partial P_{1}} = \frac{1}{2} = 0.5\ \text{Pa/Pa}
|
0.5 Pa/Pa |
| Downstream Pressure Coefficient (cP2) |
\frac{\partial\overline{P}}{\partial P_{2}} = \frac{1}{2} = 0.5\ \text{Pa/Pa}
|
0.5 Pa/Pa |
Step 1: Determine the input terms
Each transducer has four contributors. Two of them are shared between the transducers: both were calibrated against the same reference standard, and both are AutoZeroed against the same barometer. The other two are independent, because each transducer has its own drift and resolution.
The table below provides a summary of the uncertainties, sensitivity coefficients, and correlation coefficients.
| Contributor | Value | Distribution | Divisor | u(xi) | ci | r |
|---|---|---|---|---|---|---|
| P1 Reference Standard Uncertainty
(0.0016 % rdg + 0.4 Pa at 500 kPa) |
8.4 Pa | Normal | 2 | 4.2 Pa | 0.5 Pa/Pa | +1 |
| P1 AutoZero Reference
(0.005 % of 700 kPa) |
35.0 Pa | Normal | 2 | 17.5 Pa | 0.5 Pa/Pa | +1 |
| P1 Drift
(0.02 % of 700 kPa) |
140.0 Pa | Normal | 2 | 70.0 Pa | 0.5 Pa/Pa | 0 |
| P1 Resolution
(A700K) |
0.7 Pa | Rectangular | 2√3 | 0.20 Pa | 0.5 Pa/Pa | 0 |
| P2 Reference Standard Uncertainty
(0.0016 % rdg + 0.4 Pa at 500 kPa) |
8.4 Pa | Normal | 2 | 4.2 Pa | 0.5 Pa/Pa | +1 |
| P2 AutoZero Reference
(0.005 % of 700 kPa) |
35.0 Pa | Normal | 2 | 17.5 Pa | 0.5 Pa/Pa | +1 |
| P2 Drift
(0.02 % of 700 kPa) |
140.0 Pa | Normal | 2 | 70.0 Pa | 0.5 Pa/Pa | 0 |
| P2 Resolution
(A700K) |
0.7 Pa | Rectangular | 2√3 | 0.20 Pa | 0.5 Pa/Pa | 0 |
Note 1: Sensitivity coefficients are ½ because P̄ is the average of two readings.
Note 2: The 35 Pa uncertainty applies only when AutoZero isn’t used regularly.
Note 3: The 140 Pa Drift is based on the OEM calibration tolerance limits. It covers drift over the calibration interval, so no separate stability term is needed.
Step 2: Calculate the total variance
The input terms were multiplied by their respective sensitivity coefficients cito calculate the uncertainty component ui(y). Next, each uncertainty was squared to convert it to a variance ui2(y). Then, the variances were added together to calculate the total variance.
A summary is provided in the table below.
| Contributor | ci • u(xi) | [ci • u(xi)]2 |
|---|---|---|
| P1 Reference Standard Uncertainty | 2.1 Pa | 4.41 Pa² |
| P1 AutoZero Reference | 8.75 Pa | 76.56 Pa² |
| P1 Drift | 35 Pa | 1225 Pa² |
| P1 Resolution | 0.1 Pa | 0.01 Pa² |
| P2 Reference Standard Uncertainty | 2.1 Pa | 4.41 Pa² |
| P2 AutoZero Reference | 8.75 Pa | 76.56 Pa² |
| P2 Drift | 35 Pa | 1225 Pa² |
| P2 Resolution | 0.1 Pa | 0.01 Pa² |
| Total Variance | 2611.97 Pa² | |
Step 3: Calculate the total covariance
The total covariance term is calculated by multiplying, for each correlated pair, the two sensitivity coefficients ci, the two standard uncertainties u(xi), and their correlation coefficient r.
Only the two shared contributors form correlated pairs. The independent pairs have r = 0 and drop out.
| Correlated Pair | c1 • c2 • u(x1) • u(x2) • r | Covariance |
|---|---|---|
| Reference Standard (P1,P2) | ½ · ½ · 4.2 · 4.2 · 1 | 4.41 Pa² |
| AutoZero Reference (P1,P2) | ½ · ½ · 17.5 · 17.5 · 1 | 76.56 Pa² |
| Total Covariance | 80.97 Pa² | |
Step 4: Calculate the combined standard uncertainty
Finally, calculate the combined standard uncertainty using the previously calculated variance and covariance terms.
Step-by-step instructions and a summary table are provided below for reference.
- Multiply the covariance by 2.
- Add the total variance and twice the total covariance terms to calculate combined variance.
- Calculate the square root of the combined variance.
| Step | Result |
|---|---|
| 1) Multiply the covariance by 2 | 2 × 80.97 Pa²
= 161.95 Pa² |
| 2) Calculate the Combined Variance | 2611.97 Pa² + 161.95 Pa²
= 2773.91 Pa² |
| 3) Calculate the square root | √2773.91 Pa²
= 52.67 Pa |
| Combined Standard Uncertainty | 53 Pa |
Optional: Convert to Relative Uncertainty
If relative uncertainties are easier to understand, the combined standard uncertainty can (optionally) be converted to a percentage using the formula below.
A summary is provided in the table below.
| Step | Result |
|---|---|
| 1) Result (P̄) | 500 kPa |
| 2) Combined Standard Uncertainty | 52.67 Pa |
| 3) Convert Absolute to Relative Uncertainty | 100 x 52.67 Pa / 500 kPa |
| Relative Combined Standard Uncertainty | 0.011 % |
Compare the results
The table below compares results based on the assigned correlation coefficients.
- When all contributors are set to r = 0, the combined standard uncertainty is slightly understated.
- When all contributors are set to r = +1, the combined standard uncertainty is significantly overstated.
Make sure to assign r from shared error sources, not by default. Otherwise, the uncertainty could be overstated or understated.
| Assumption | uc(P̄) | Comparison |
|---|---|---|
| All contributors uncorrelated (r = 0) | 51.11 Pa | 3.0 % understated |
| Shared contributors correlated (correct) | 52.67 Pa | — |
| All contributors fully correlated (r = +1 for each P₁/P₂ pair) | 72.28 Pa | 37.2 % overstated |
Note: The percentages (%) for overstated and understated are relative to the correct result (52.67 Pa).
When the combined standard uncertainty is rounded to two significant digits, the correct result is 53 Pa and the uncorrelated result (r = 0) is 51 Pa. Because the reported value changes, the correlation is significant here, even though it is small.
Takeaway: Averaging two transducers reduces only the independent part of their uncertainty. The shared reference and AutoZero contributions pass through the average unchanged.
Compare this with the ISO 16063-21 example below. Both share a reference standard (r = +1), but an averaging formula carries the shared error through, while a ratio formula cancels it.
Correlated Example 2: Voltage Ratio Measurement (ISO 16063-21, Calibration by Comparison)
ISO 16063-21, Table A.1 includes uncertainty components for the determination of accelerometer sensitivity magnitude. Row 3 in the table specifies “Voltage ratio measurement (often two correlated measurements)” as a source of uncertainty, making it a great example for this guide.
The ISO 16063-21 standard presents this uncertainty as one contributor. However, it is actually the combination of multiple contributors, such as reference standard uncertainty, drift, resolution, and more.
For simplicity, this example will only evaluate the drift and reference standard uncertainty of the equipment used to measure the voltage ratio.
Measurement Function
In accelerometer calibration by back-to-back comparison, the UUT magnitude sensitivity is determined from the ratio of the two transducer voltages and the reference transducer sensitivity. The formula is provided below for reference.
| Formula | Input Terms |
|---|---|
|
S_{UUT} = S_{Ref} \bullet \frac{E_{UUT}}{E_{Ref}}
|
SRef = Reference Transducer magnitude sensitivity
ERef = Reference Transducer output voltage SUUT = UUT Transducer magnitude sensitivity EUUT = UUT Transducer output voltage |
The table below gives the input quantities for this example.
| Quantity | Value |
|---|---|
| Acceleration Level | 10 g at 100 Hz |
| Reference Transducer Sensitivity (SRef) | 98.8 mV/g |
| Reference Transducer Voltage (ERef) | 988 mV |
| UUT Transducer Voltage (EUUT) | 100 mV |
| Result: UUT Transducer Sensitivity (SUUT) | 98.8 x (100/988) = 10.0 mV/g |
Sensitivity Coefficients
In the table below, sensitivity coefficients are determined from the measurement model for the channel 1 and channel 2 terms.
| Coefficients | Formula | Sensitivity Coefficient |
|---|---|---|
| Channel 1 Sensitivity Coefficient (cCh1) |
\frac{\partial S_{UUT}}{\partial E_{Ref}} = \frac{{- S}_{Ref} \bullet E_{UUT}}{E_{Ref}^{2}}
|
-0.010121 (mV/g)/mV |
| Channel 2 Sensitivity Coefficient (cCh2) |
\frac{\partial S_{UUT}}{\partial E_{UUT}} = \frac{S_{Ref}}{E_{Ref}}
|
0.1 (mV/g)/mV |
Which Contributors Are Correlated?
The VR9700 controller measures the transducer voltages on different channels. Each channel drifts independently, so the drift contributors are uncorrelated (r = 0). However, both channels were calibrated against the same reference standard, so the reference standard contributors are correlated (r = +1).
Therefore, the uncertainties are combined using the correlated input quantity formula from JCGM 100:2008, § 5.2.2.
| Contributor | Correlation |
|---|---|
| Drift | Independent, r = 0 |
| Reference Standard Uncertainty | Shared, r = +1 |
Step 1: Determine the input terms
This evaluation includes four uncertainty contributors, including the drift and reference standard uncertainty for each channel (i.e., Channel 1 and 2).
- The drift uncertainties were characterized with a rectangular distribution.
- The reference standard uncertainties were characterized with a normal distribution.
- The sensitivity coefficients were determined from the measurement function and their signs (+,-) were kept.
The formula and table below show how the relative uncertainties are converted to absolute uncertainties.
| Contributor | Input Value
xi |
Relative Half-Width
arel,i |
Absolute Half-Width
ai |
|---|---|---|---|
| Ch 1 Drift | 988 mV | 0.5 % | 4.94 mV |
| Ch 1 Reference Standard | 988 mV | 0.1 % (k=2) | 0.988 mV (k=2) |
| Ch 2 Drift | 100 mV | 0.5 % | 0.500 mV |
| Ch 2 Reference Standard | 100 mV | 0.1 % (k=2) | 0.100 mV (k=2) |
The table below provides a summary of the uncertainties, sensitivity coefficients, and correlation coefficients.
| Contributor | ai
(mV) |
Distribution | Divisor | u(xi)
(mV) |
ci
(mV/g)/mV |
r |
|---|---|---|---|---|---|---|
| Ch 1 Drift | 4.94 | Rectangular | √3 | 2.852 | -0.010121 | 0 |
| Ch 1 Reference Standard | 0.988 | Normal | 2 | 0.494 | -0.010121 | +1 |
| Ch 2 Drift | 0.500 | Rectangular | √3 | 0.289 | 0.10000 | 0 |
| Ch 2 Reference Standard | 0.100 | Normal | 2 | 0.050 | 0.10000 | +1 |
Note: The negative sign on cCh1 is the key to this example. A higher reference voltage gives a lower UUT sensitivity.
Step 2: Calculate the total variance
The input terms were multiplied by their respective sensitivity coefficients ci to calculate the uncertainty component ui(y). Next, each uncertainty was squared to convert it to a variance ui2(y). Then, the variances were added together to calculate the total variance.
A summary is provided in the table below.
| Contributor | ci • u(xi) | [ci • u(xi)]2 |
|---|---|---|
| Ch 1 Drift (ERef) | -0.0289 mV/g | 0.000833 (mV/g)2 |
| Ch 1 Reference Standard (ERef) | -0.0050 mV/g | 0.000025 (mV/g)2 |
| Ch 2 Drift (EUUT) | +0.0289 mV/g | 0.000833 (mV/g)2 |
| Ch 2 Reference Standard (EUUT) | +0.0050 mV/g | 0.000025 (mV/g)2 |
| Total Variance | 0.001717 (mV/g)2 | |
Step 3: Calculate the total covariance
The total covariance term is calculated by multiplying, for each correlated pair, the two sensitivity coefficients ci, the two standard uncertainties u(xi), and their correlation coefficient r.
The formula and calculation are provided below for reference.
| Contributor | cCh1 • cCh2 • u(ERef) • u(EUUT) • r | Covariance |
|---|---|---|
| Ch 1 & Ch 2 Drift | -0.010121· 0.1 · 2.852 · 0.289 · 0 | 0.000000 (mV/g)² |
| Ch 1 & Ch 2 Reference Standard | -0.010121· 0.1 · 0.494 · 0.05 · +1 | -0.000025 (mV/g)² |
| Total Covariance | -0.000025 (mV/g)² | |
Step 4: Calculate the combined standard uncertainty
Finally, calculate the combined standard uncertainty using the previously calculated variance and covariance terms.
Step-by-step instructions and a summary table are provided below for reference.
- Multiply the covariance by 2.
- Add the total variance and twice the total covariance terms to calculate combined variance.
- Calculate the square root of the combined variance.
| Step | Result |
|---|---|
| 1) Multiply the covariance by 2 | 2 × -0.000025 (mV/g)²
= -0.000050 (mV/g)² |
| 2) Calculate the Combined Variance | 0.001717 (mV/g)² – 0.000050 (mV/g)²
= 0.001667 (mV/g)² |
| 3) Calculate the square root | √0.001667 (mV/g)²
= 0.0408 mV/g |
| Combined Standard Uncertainty | 0.041 mV/g |
Optional: Convert to Relative Uncertainty
If relative uncertainties are easier to understand, the combined standard uncertainty can (optionally) be converted to a percentage using the formula below.
A summary is provided in the table below.
| Step | Result |
|---|---|
| 1) Result (SUUT) | 10.0 mV/g |
| 2) Combined Standard Uncertainty | 0.0408 mV/g |
| 3) Convert Absolute to Relative Uncertainty | 100 x 0.0408 mV/g / 10.0 mV/g |
| Relative Combined Standard Uncertainty | 0.41 % |
Compare the results
The table below compares results based on the assigned correlation coefficients.
- When all contributors are set to r = 0, the combined standard uncertainty is slightly overstated.
- When all contributors are set to r = +1, the combined standard uncertainty is significantly understated.
Make sure to assign r from shared error sources, not by default. Otherwise, the uncertainty could be overstated or understated.
| Assumption | uc,rel(SUUT) | Comparison |
|---|---|---|
| All contributors uncorrelated (r = 0) | 0.414 % | 1.5 % (relative) overstated |
| Shared contributors correlated (correct) | 0.408 % | — |
| All contributors fully correlated (r = +1 for each Ch 1/Ch 2 pair) | 0.000 % | 100 % understated |
Both uncertainties, 0.414 % and 0.408 %, round to 0.41 %, so the correlation here is real but not significant (see the significance test above).
Takeaway: In a ratio, the shared reference standard error cancels because the coefficients have opposite signs. The independent drift remains.
How to Calculate Combined Standard Uncertainty in Excel
Calculate the combined standard uncertainty in Excel with one of three formulas, depending on whether your budget includes sensitivity coefficients or correlated inputs. The table below summarizes them, and the examples below show which cells to select. For a full walkthrough, see how to create an uncertainty budget in Excel.
Other techniques, such as a correlation matrix (see the FAQ), also work. Whichever you use, validate your spreadsheet against a known result before relying on it.
Summary of Excel Functions for Combined Standard Uncertainty
| Conditions | Excel Function |
|---|---|
| No sensitivity coefficients | =SQRT(SUMSQ([u(xi) cells])) |
| With sensitivity coefficients | =SQRT(SUMPRODUCT([ci cells]^2,[u(xi) cells]^2)) |
| Correlated input quantities | =SQRT(SUMPRODUCT([ci cells]^2,[u(xi) cells]^2)+2*SUM([covariance cells])) |
Note: Excel calculates with full precision, so values shown in the images are rounded for display. Your recalculated results may differ slightly.
Example 1: No Sensitivity Coefficients
This example shows how to calculate the combined standard uncertainty when no sensitivity coefficients or covariances are involved.
The formula below calculates the square root of the sum of squares for all selected standard uncertainty cells. Use this formula when your calculator doesn’t use sensitivity coefficients, meaning every uncertainty is already in the same unit as the result.
Copy and paste this formula into Excel and replace (including the brackets) [u(xi) cells] with the cell range containing the standard uncertainties. The image below shows an example of using this function and which cells to select.
The result from the above image is uc(y) = 0.006044 psig, rounded to 0.0060 psig.
Example 2: With Sensitivity Coefficients
This example shows how to calculate the combined standard uncertainty when sensitivity coefficients are involved with no covariances.
The formula below calculates the sum of squared products of all selected sensitivity coefficients and standard uncertainties, then calculates the square root of the total.
Copy and paste this formula into Excel, and replace (including the brackets):
- [ci cells] with the cell range containing the sensitivity coefficients, and
- [u(xi) cells] with the cell range containing the standard uncertainties.
The image below shows an example of using this function and which cells to select.
The result from the above image is uc(y) = 0.006044 psig, rounded to 0.0060 psig.
Example 3: With Correlated Input Quantities
This example shows how to calculate the combined standard uncertainty when sensitivity coefficients and covariances are involved.
The formula below calculates the sum of squared products of all selected sensitivity coefficients and standard uncertainties plus twice the sum of the covariances, then calculates the square root of the total.
Copy and paste this formula into Excel and replace (including the brackets):
- [ci cells] with the cell range containing the sensitivity coefficients,
- [u(xi) cells] with the cell range containing the standard uncertainties, and
- [covariance cells] with the cell range containing the covariances.
The image below shows an example of using this function and which cells to select. This is for illustrative purposes only and may not represent the typical format in which covariances are presented. Most uncertainty calculators evaluate and present covariances in a separate table.
Note: Use the signed sensitivity coefficients (+ or −) in each covariance cell. A negative coefficient makes the covariance negative and reduces the combined standard uncertainty (see Correlated Example 2).
The result from the above image is uc(y) = 0.006083 psig, rounded to 0.0061 psig.
Common Mistakes When Calculating Combined Standard Uncertainty
Most errors in combined standard uncertainty come from a few repeated mistakes. Use the table below to check your calculations.
| Mistake | Why It’s Wrong | How to Fix It |
|---|---|---|
| Before combining | ||
| Combining expanded uncertainties (U) | Expanded uncertainties have different coverage factors and overstate the result. | Divide each U by its coverage factor (k) to get the standard uncertainty first. |
| Using a half-width as a standard uncertainty | Skipping the divisor overstates the contributor. | Divide by the divisor for its distribution (e.g., √3 for rectangular). |
| Combining contributors in different units | Volts, ohms and amperes can’t be added together. | Convert each contributor to the unit of the result using sensitivity coefficients or relative uncertainties. |
| Assuming every sensitivity coefficient is 1 | The model decides the coefficient, not the units. Averaging two readings gives c = ½. | Determine each coefficient from the measurement function. |
| Rounding intermediate values | Early rounding builds up error. Exact cancellations no longer reach zero. | Keep full precision until the final result, then round to two significant digits. |
| Correlation | ||
| Treating all inputs as uncorrelated by default | Shared error sources are ignored, which can understate or overstate the result. | Check for shared sources: the same reference standard, instrument, or environment. |
| Applying r = +1 to an entire specification | Only part of a specification is usually shared. The result can cancel to zero. | Split the specification into shared (r = +1) and independent (r = 0) contributors. |
| Assuming averaging always reduces uncertainty | Averaging reduces only the independent part. Shared errors pass through unchanged. | Model shared contributors as correlated before averaging. |
| Covariance | ||
| Dropping the signs of sensitivity coefficients | Negative coefficients make a covariance term negative. Without the sign, nothing cancels. | Keep each coefficient’s sign (+ or −) in the covariance terms. |
| Applying sensitivity coefficients twice | Using ci·u(xi) values and then multiplying by ci again shrinks the covariance. | Multiply each standard uncertainty u(xi) by its coefficient only once. |
| Using the wrong uncertainty in a covariance term | A covariance term built from another pair’s values gives a wrong result. | Use the two standard uncertainties from the same correlated pair. |
| Counting a pair twice, or forgetting the factor of 2 | The covariance is doubled twice or not at all. | Count each pair once (i < j), then multiply the total covariance by 2. |
| Subtracting a positive covariance | Positive coefficients with r = +1 increase the uncertainty, not decrease it. | Add twice the total covariance. Its sign comes from the coefficients and r. |
| Results | ||
| Trusting a zero result | Zero means the shared error cancels, not that the measurement has no uncertainty. | Check which contributors cancelled and confirm the independent ones remain. |
| Reporting uc on a certificate | Accredited calibration certificates require an expanded uncertainty with its coverage factor. | Convert uc to an expanded uncertainty first. |
| Writing “ppm” without units | Parts-per notations are not SI units and are ambiguous. | Write the unit ratio, such as µA/A (SI Brochure § 5.4.7; NIST SP 811 § 7.10.3). |
Quick checks for correlated calculations
Below is a list of quick checks that should be made for all correlated calculations to ensure uncertainties are accurate.
- All r = 0: uc should equal the root-sum-square of the ci·u(xi) values.
- All r = +1: uc should equal the absolute value of the linear sum, |Σ ci·u(xi)|.
- If your result doesn’t match these checks, a covariance term is using the wrong values or the wrong sign.
Conclusion
Calculate combined standard uncertainty by converting every contributor to a standard uncertainty in the unit of the result and combining them with the root sum of squares. When inputs share an error source, add the covariance terms. The method stays the same no matter how many contributors your budget has. What changes the result is how carefully you set up the inputs.
Before you finalize your combined standard uncertainty, check that you have:
- Converted every contributor to a standard uncertainty using the correct distribution and divisor.
- Applied each sensitivity coefficient once, keeping its sign (+ or −).
- Checked for shared error sources, such as the same reference standard, instrument, or environment.
- Assigned r = +1 only to the shared part of a specification, not the whole specification.
- Confirmed the result with the quick checks (all r = 0 and all r = +1).
- Kept full precision until the final result.
The combined standard uncertainty isn’t the final step. To report your uncertainty on a calibration certificate or a scope of accreditation, convert it to an expanded uncertainty with a coverage factor. Learn how in How to Calculate Expanded Uncertainty.
To save time and avoid calculation errors, try the Simple Uncertainty Calculator used in this guide. If you’d like help building or reviewing your uncertainty budgets, contact me.
Frequently Asked Questions
Below are answers to frequently asked questions we receive from lab owners, managers, and technicians.
Basics
What is combined standard uncertainty?
Combined standard uncertainty, uc(y), is the standard uncertainty of a measurement result. It is calculated by combining the standard uncertainties of all input quantities in the measurement model, using the root sum of squares plus covariance terms when inputs are correlated (JCGM 200:2012, § 2.31; JCGM 100:2008, § 5).
What is the formula for combined standard uncertainty?
There are two formulas for calculating the combined standard uncertainty. One for uncorrelated input quantities and one for correlated input quantities. Both are provided in the table below.
What is the symbol for combined standard uncertainty?
The symbol used for combined standard uncertainty is a lower-case u with a subscript c, uc.
What is the difference between combined and expanded uncertainty?
Combined standard uncertainty (uc) is a standard uncertainty, a standard deviation equivalent. For a normal distribution, it corresponds to a level of confidence of approximately 68 %. Expanded uncertainty (U) is the combined standard uncertainty multiplied by a coverage factor (k) greater than one, to give a higher level of confidence.
Typically, expanded uncertainties use k = 2, which gives approximately 95 % (95.45 %) for a normal distribution.
The table below shows the key differences between combined and expanded uncertainty.
| Criteria | Combined Uncertainty | Expanded Uncertainty |
|---|---|---|
| Formula |
u_{c}(y) = \sqrt{\sum_{i = 1}^{n}\left\lbrack c_{i}u\left( x_{i} \right) \right\rbrack^{2}}
|
U = k \bullet u_{c}(y)
|
| VIM Definition
(JCGM 200:2012) |
2.31 – standard measurement uncertainty that is obtained using the individual standard measurement uncertainties associated with the input quantities in a measurement model. | 2.35 – the product of a combined standard measurement uncertainty and a factor larger than the number one. |
| Level of confidence | ≈ 68 % (normal distribution) | > 68 %
(Typically, 95.45 %) |
| Coverage Factor | k = 1 | k > 1
(Typically, k = 2) |
| Reported in Accredited Reports | No (calibration certificates require U) | Yes |
Is standard uncertainty the same as standard deviation?
Not exactly. A standard uncertainty is measurement uncertainty expressed as a standard deviation (JCGM 200:2012, § 2.30). A Type A standard uncertainty comes from the standard deviation of measured data, while a Type B standard uncertainty is the standard deviation of an assumed probability distribution. So, it is a standard deviation equivalent, not always a standard deviation of data.
Is standard uncertainty the same as standard error?
No, standard uncertainty and standard error are not the same, although they are closely related and often confused.
Standard error (SE) is the standard deviation of the sampling distribution of a statistic, such as the sample mean. It describes how much an estimate is expected to vary from sample to sample, and it is used in statistics and data analysis.
Standard uncertainty is measurement uncertainty expressed as a standard deviation (JCGM 200:2012, § 2.30). It can be evaluated statistically (Type A) or by other means, such as calibration data or scientific judgment (Type B). It is used in metrology to express the doubt associated with a measurement result.
The two overlap in one case: the Type A standard uncertainty of a mean, s/√n, is numerically the same as the standard error of the mean.
Method
Why do you use the root sum of squares to combine uncertainties?
The root sum of squares, also called adding in quadrature, is used because variances of independent contributors add, but their standard deviations do not.
You cannot combine standard deviations through linear addition. They must be converted to variances (i.e., squared) first before adding them together. Once the combined variance is determined, the square root of the variance is calculated to convert it back to a standard deviation. Hence the phrase root sum of squares – the square root of the sum of squares.
Like the Pythagorean theorem and vector addition, independent contributors combine like the perpendicular sides of a right triangle: the combined uncertainty is the total magnitude or length of the triangle’s hypotenuse.
Each contributor also has its own probability distribution. When many are combined, the resulting distribution tends to be approximately normal (Central Limit Theorem), as long as certain conditions are met, which they typically are. See JCGM 100:2008, § G.2 and § G.6.6.
Can I add uncertainties linearly instead of using RSS?
Yes, but only for contributors that are fully correlated (r = +1) with sensitivity coefficients of the same sign. In that case, the combined uncertainty equals their linear sum. Adding independent uncertainties linearly overstates the combined uncertainty and is not the GUM method.
For example, OIML R111, § C.6.2.2 states that “If a combination of reference weights is used for a mass comparison and their covariances are not known, a correlation coefficient of 1 can be assumed. This will lead to linear summation of uncertainties:”
The same approach applies when gauge blocks calibrated against the same reference are wrung together, and in other cases where contributors share a common source of uncertainty.
How do you combine uncertainties with different units?
Uncertainty contributors with different units of measurement can be combined using one of the following techniques:
- Propagate each uncertainty through the measurement function (one factor at a time) – Change one input by its standard uncertainty, recalculate the result, and record the change. For nonlinear functions, change the input up and down, and use half the difference.
- Propagate each uncertainty using propagation rules – Use propagation rules based on the measurement function to convert each uncertainty to the units of the result (see John R. Taylor, An Introduction to Error Analysis).
- Use sensitivity coefficients – Determine each coefficient from the measurement function, then multiply each standard uncertainty by its coefficient to convert it to the same units as the result.
- Use relative uncertainties – Convert each uncertainty to a ratio or percentage of its input. This works when the measurement function is a product or quotient, but not for all measurement functions.
How do you combine uncertainties when multiplying or dividing?
Express each uncertainty relative to its input (e.g., as a ratio or percentage), then combine them using the root sum of squares. For a product or quotient of uncorrelated inputs, the relative uncertainty of the result is the root sum of squares of the inputs’ relative uncertainties.
John R. Taylor’s An Introduction to Error Analysis gives the following propagation rules for products and quotients of uncorrelated inputs.
| Function | Propagated Uncertainty |
|---|---|
|
z = cxy
|
\frac{\delta z}{z} = \left\lbrack \left( \frac{\delta x}{x} \right)^{2} + \left( \frac{\delta y}{y} \right)^{2} \right\rbrack^{\frac{1}{2}}
|
|
z = c\frac{x}{y}
|
\frac{\delta z}{z} = \left\lbrack \left( \frac{\delta x}{x} \right)^{2} + \left( \frac{\delta y}{y} \right)^{2} \right\rbrack^{\frac{1}{2}}
|
JCGM 100:2008, § 5.1.6 gives the same relative form for products and powers, where pi is the exponent of each input:
How do you calculate a sensitivity coefficient?
A sensitivity coefficient (ci) is the partial derivative of the measurement function with respect to an input quantity. In practice, it is the change in the result divided by the change in the input, keeping the units.
For example, a 1-inch steel gauge block with a thermal expansion coefficient of 10.8 × 10⁻⁶/°C (i.e., Mitutoyo steel gauge block) changes length by 10.8 µin for each 1 °C change in temperature. The sensitivity coefficient is 10.8 µin/°C (or 10.8 µin/K).
Whenever possible, determine sensitivity coefficients from the measurement function. Otherwise, determine them experimentally.
What is relative combined standard uncertainty?
Relative combined standard uncertainty is the combined standard uncertainty divided by the absolute value of the result, uc(y)/|y| (for y ≠ 0). It can be expressed as:
- Ratio: unit per unit (e.g., Ω/Ω)
- Percentage: percentage of the result (e.g., 100 × ratio).
- Parts per million: written with the units of the result (e.g., µΩ/Ω = 10⁶ × ratio).
How do you calculate combined uncertainty in Excel?
Calculate combined uncertainty in Excel by combining the SQRT and SUMSQ functions. If the sensitivity coefficients are in a separate column, use SQRT with SUMPRODUCT.
| Conditions | Excel Function |
|---|---|
| No sensitivity coefficients | =SQRT(SUMSQ([u(xi) cells])) |
| With sensitivity coefficients | =SQRT(SUMPRODUCT([ci cells]^2,[u(xi) cells]^2)) |
| Correlated input quantities | =SQRT(SUMPRODUCT([ci cells]^2,[u(xi) cells]^2)+2*SUM([covariance cells])) |
Note 1: Replace each placeholder, including the brackets, with the matching cell range.
Note 2: Calculate each covariance cell as [ci·u(xi)] × [cj·u(xj)] × r for each correlated pair, keeping the signs.
Alternative Technique: For correlated inputs, use a correlation matrix:
=SQRT(SUMPRODUCT(MMULT([r matrix],[ci·u(xi) cells]),[ci·u(xi) cells]))
The ci·u(xi) values (with their signs) are in one column, and the r matrix has 1s on its diagonal. In older Excel versions, enter it as an array formula (Ctrl+Shift+Enter).
Correlation
What is the difference between correlated and uncorrelated uncertainties?
Uncorrelated uncertainties come from independent error sources, so their correlation coefficient is zero (r = 0). Correlated uncertainties share an error source, such as the same reference standard, instrument, or environmental influence, so an error in one also appears in the other (r ≠ 0).
How do you combine correlated uncertainties?
To combine correlated uncertainties, calculate the total variance of all input terms and the covariance of each correlated pair. Then add the total variance to twice the total covariance, and take the square root to get the combined standard uncertainty.
Refer to JCGM 100:2008, § 5.2.2 for more information.
What correlation coefficient counts as significant?
JCGM 100:2008 does not set a threshold for a significant correlation. To decide, perform an A/B test – calculate the combined standard uncertainty both ways (below) and compare:
- A: Evaluate uncertainty with correlation coefficients.
- B: Evaluate uncertainty without correlation coefficients.
If the difference doesn’t change your reported uncertainty after rounding to two significant digits, the correlation is not significant.
Can uncertainty be negative?
No. JCGM 200:2012, § 2.26 defines measurement uncertainty as a “non-negative parameter.”
Individual covariance terms can be negative. That happens when inputs share an error source and their sensitivity coefficients have opposite signs (see Correlated Example 2). But the combined variance can’t be negative when the correlation coefficients are valid. A negative combined variance means there is a calculation error or inconsistent correlation coefficients.
Does averaging two instruments reduce uncertainty?
Yes, but only partly. Averaging n instruments reduces their independent uncertainties by about 1/√n, because each sensitivity coefficient is 1/n. Shared uncertainties, such as the same reference standard, pass through the average unchanged.
See Correlated Example 1 (Fluke molbox) earlier in this guide.
Practical Questions
Should I add the reference standard uncertainty if I’m already using the manufacturer’s accuracy specification?
It depends on whether the accuracy specification already includes the reference standard uncertainty. If it doesn’t, add the reference standard uncertainty as a separate term. If it does, adding it again double-counts it (JCGM 100:2008, § 4.3.10), unless the equipment was calibrated by someone other than the OEM or under different conditions. The common cases are below.
General Accuracy Specifications — Typically, the accuracy specification does not include the reference standard (i.e., calibration) uncertainty in the specification. In this case, the reference standard uncertainty must be included as an independent term in the uncertainty analysis.
Absolute Accuracy Specifications with OEM Calibrations — If the accuracy specification is claimed to be an absolute accuracy specification, such as many Fluke specifications, then the reference standard uncertainty is included in the accuracy specification and including an additional reference standard uncertainty input term will result in double-counting uncertainty so long as the equipment is calibrated by the OEM under conditions defined by the specifications. Double-counting uncertainty is discouraged by JCGM 100:2008, § 4.3.10.
Absolute Accuracy Specifications with non-OEM Calibrations — If equipment is calibrated by a supplier other than the OEM, then the reference standard uncertainty included in the absolute accuracy specification is invalid and must be replaced with the actual reference standard uncertainty. When the OEM reference standard uncertainty is unknown (typical), the reference standard uncertainty from the non-OEM supplier must be included as an independent term in the uncertainty analysis.
Absolute Accuracy Specifications with OEM Calibrations under Different Conditions — If the OEM calibrates the equipment under conditions that don’t match the published specifications, the reference standard portion of the absolute specification no longer applies. Include the reference standard uncertainty from the calibration certificate as a separate term.
Accuracy Specifications with Traceability Adders — If the specification is stated relative to calibration standards and gives a separate traceability adder (e.g., Keysight 3458A), don’t use the adder value. Instead, include the actual reference standard uncertainty from the calibration report as a separate term.
Should I use the calibration tolerance or the manufacturer’s specification?
Use the calibration tolerance when it is larger than the manufacturer’s specification. When the calibration tolerance is tighter, use it only if you can justify it, and confirm it isn’t an adjustment threshold. Otherwise, use the manufacturer’s specification. The common cases are below.
Calibration Tolerance > Manufacturer Specification
If the calibration tolerance is greater than the manufacturer specification, use the calibration tolerance. Manufacturer specifications can change without notice, and OEM documents aren’t always kept up to date. An OEM calibration report can show this if the calibration tolerance has increased since the last calibration and is larger than the published specification. External calibration providers may also calibrate to the wrong specification or tolerance. In either case, use the calibration tolerance, because the laboratory can only guarantee that the equipment conforms to those limits. To use the manufacturer’s specification instead, the lab must technically justify it and support it with objective evidence.
Manufacturer Specification > Calibration Tolerance: OEM Calibration
If the calibration tolerance is less than the manufacturer specification and the calibration was performed by the OEM, then the calibration tolerance is acceptable for use. However, the lab should contact the OEM to confirm the reported tolerance limits are correct and obtain the reason for the change (i.e., reduction). Additionally, the lab should confirm that the tighter tolerances are not adjustment thresholds applicable to as-found calibration data. It is common for labs to use tighter limits to trigger adjustment actions and these should not be used in place of manufacturer specifications or full calibration tolerance limits.
Manufacturer Specification > Calibration Tolerance: Non-OEM Calibration
If the calibration tolerance is less than the manufacturer specification and the calibration was performed by an external calibration provider, then the manufacturer specification should be used unless the lab can technically justify that equipment has consistently performed within the tighter calibration tolerance limit (e.g., over the last 3 to 5 calibrations).
Adjustment Thresholds
If the calibration report shows a tolerance tighter than the manufacturer specification, confirm it isn’t an adjustment threshold. Many labs, both OEM and non-OEM, use tighter limits to decide when to adjust, and those limits don’t define conformance. Use the tolerance that the instrument is declared to conform to (the pass/fail limit). Labels vary between labs. Many apply the adjustment threshold to as-found data and adjust any point that exceeds it. Others list it with the as-left data. For example, a Fluke molbox1 calibration report shows a 0.140 kPa as-found specification and a 0.070 kPa as-left adjustment specification.
90-Day Specifications
If equipment is calibrated to 90-day specifications (e.g., Fluke) but the equipment is calibrated annually, then the lab should use the 1-year specification. If equipment is calibrated quarterly, then the 90-day specification is acceptable for use.
Environmental Conditions
If equipment is calibrated under environmental conditions that don’t match the manufacturer specifications used in the uncertainty analysis (e.g., ±1 °C vs. ±5 °C specifications), use the manufacturer specifications that match the calibration environment. If possible, apply corrections to normalize results to the environmental conditions of the desired specification.
Can my uncertainty be smaller than my reference standard’s uncertainty?
Usually not. When the reference standard contributes directly to the result (c = 1), your combined and expanded uncertainties should be at least as large as the reference standard’s uncertainty at the same level of confidence. The exceptions are measurements where the reference’s sensitivity coefficient is less than 1 (e.g., averaging) or where its contribution cancels (e.g., a ratio; see Correlated Example 2).
If your uncertainty is smaller in a direct comparison, check for these common mistakes:
- Wrong uncertainty value – using the wrong reference standard uncertainty value can cause the uncertainty to be smaller than the reference standard uncertainty.
- Miscalculation of uncertainty value – using the wrong reference standard uncertainty formula1 or the wrong sign (i.e., +,-) can cause the uncertainty to be smaller than the reference standard uncertainty.
- Wrong distribution – using the wrong probability distribution can cause the uncertainty to be smaller than the reference standard uncertainty.
- Wrong divisor – using the wrong divisor can cause the uncertainty to be smaller than the reference standard uncertainty.
- Wrong combined uncertainty formula – using the wrong formula to calculate the combined standard uncertainty can cause the uncertainty to be smaller than the reference standard uncertainty. (e.g., correlated vs uncorrelated formulas)
- Wrong coverage factor – using the wrong coverage factor can cause the uncertainty to be smaller than the reference standard uncertainty (e.g., k = 2 vs k = 2.4).
Note 1: Accredited labs are not supposed to report CMC formulas in accredited reports, but it is common in many dimensional and mechanical calibration reports. It requires customers to calculate and round their own uncertainties, which is the main cause of miscalculations. Reporting CMC functions directly from the scope of accreditation violates Global ACI TECH-1-009, § 5.2 (formerly ILAC P14).
When should I use Monte Carlo instead of the GUM method?
Use the Monte Carlo method (JCGM 101:2008) when a test or calibration method requires it (e.g., ISO/TS 15530-4), or when the assumptions of the GUM method don’t hold.
The list below gives scenarios where the Monte Carlo method is preferred over the GUM method:
- Non-linear measurement functions.
- Complex partial derivatives.
- Non-normal output distributions.
- Complex or asymmetrical input distributions.
- Input quantities with strong interactions.
- Validating a result calculated with the GUM method.
To use the Monte Carlo method, you need the measurement function and each input’s value, uncertainty, and probability distribution. Errors in any of these produce misleading results. To try it, use the free NIST Uncertainty Machine, which calculates results with both the GUM and Monte Carlo methods.
Terms and Definitions
The terms below are used throughout this guide. Symbols follow JCGM 100:2008 (GUM) notation. Definitions are written in plain language; see the referenced clause for the official wording. For more terms, see uncertainty budget definitions.
- Combined standard uncertainty (uc(y))
- The standard uncertainty of the measurement result, calculated by combining the standard uncertainties of the input quantities in the measurement model (JCGM 200:2012, § 2.31; JCGM 100:2008, § 5.1.2).
- Combined variance (uc2(y))
- The sum of the squared uncertainty components, plus twice the covariance terms when inputs are correlated. Its positive square root is the combined standard uncertainty (JCGM 100:2008, § 5.1.2, § 5.2.2).
- Input quantities that share an error source, such as the same reference standard, instrument, or environmental influence. An error in one also appears in the other (JCGM 100:2008, § 5.2).
- Correlation coefficient (r(xi, xj))
- A number from −1 to +1 that describes how strongly two input quantities vary together. It equals their covariance divided by the product of their standard uncertainties. r = 0 means uncorrelated; r = +1 means fully correlated (JCGM 100:2008, § 5.2.2, § C.3.6).
- Covariance (u(xi, xj))
- A measure of how two input quantities vary together. In the combined uncertainty formula, each covariance term is ci·cj·u(xi)·u(xj)·r(xi, xj) (JCGM 100:2008, § 5.2.2, § C.3.4).
- Coverage factor (k)
- The number the combined standard uncertainty is multiplied by to get the expanded uncertainty. Typically k = 2 (JCGM 200:2012, § 2.38).
- Divisor
- The number an uncertainty is divided by to convert it to a standard uncertainty, based on its probability distribution. For example: 2 for a normal distribution stated at k = 2, √3 for a rectangular half-width, and 2√3 for a full resolution interval (JCGM 100:2008, § 4.3.3, § 4.3.7).
- Expanded uncertainty (U)
- The combined standard uncertainty multiplied by a coverage factor: U = k·uc(y). Accredited calibration certificates report U, not uc (JCGM 200:2012, § 2.35).
- Half-width (a)
- Half the width of the interval (±a) that a value is believed to lie within, such as a tolerance or specification limit (JCGM 100:2008, § 4.3.7).
- Independent
- Two input quantities whose errors do not influence each other. Independent inputs are always uncorrelated (r = 0), but uncorrelated inputs are not always independent (JCGM 100:2008, § C.3.7).
- Input estimate (xi)
- The value of an input quantity used to calculate the result, such as a meter reading or a certified value (JCGM 100:2008, § 4.1.4).
- Input quantity (Xi)
- A quantity that must be measured or known to calculate the measurand, such as voltage and resistance when calculating current (JCGM 200:2012, § 2.50).
- Measurand (Y)
- The quantity intended to be measured (JCGM 200:2012, § 2.3).
- Measurement function (f)
- The formula that calculates the result from the input quantities: y = f(x1, x2, …, xn). It is the measurement model written as an equation (JCGM 200:2012, § 2.48, § 2.49; JCGM 100:2008, § 4.1.1).
- Measurement uncertainty
- A non-negative parameter that characterizes the dispersion of the values attributed to a measurand (JCGM 200:2012, § 2.26).
- Output estimate (y)
- The calculated value of the measurand (i.e., the measurement result) (JCGM 100:2008, § 4.1.4).
- Probability distribution
- Describes how likely the possible values of a quantity are. Common distributions in uncertainty analysis are normal, rectangular, triangular, and U-shaped (JCGM 100:2008, § 4.3, § C.2.3).
- Relative combined standard uncertainty (uc,rel(y))
- The combined standard uncertainty divided by the absolute value of the result, uc(y)/|y|. It is expressed as a ratio (e.g., µA/A) or a percentage (JCGM 200:2012, § 2.32).
- Root sum of squares (RSS)
- The square root of the sum of squared values, also called adding in quadrature. It is used to combine uncorrelated uncertainty components (JCGM 100:2008, § 5.1.2).
- Sensitivity coefficient (ci)
- The partial derivative of the measurement function with respect to an input (∂f/∂xi). It shows how much the result changes per unit change in the input and converts the input’s units to the result’s units. Keep its sign (+ or −) (JCGM 100:2008, § 5.1.3).
- Standard uncertainty (u(xi))
- Measurement uncertainty expressed as a standard deviation (JCGM 200:2012, § 2.30).
- Type A evaluation
- Evaluation of uncertainty by statistical analysis of measured values, such as the standard deviation of repeated readings (JCGM 200:2012, § 2.28; JCGM 100:2008, § 4.2).
- Type B evaluation
- Evaluation of uncertainty by other means, such as calibration certificates, manufacturer specifications, resolution, or experience (JCGM 200:2012, § 2.29; JCGM 100:2008, § 4.3).
- Uncertainty budget
- A statement of the uncertainty contributors, with how each was evaluated and combined (JCGM 200:2012, § 2.33).
- Uncertainty component (ui(y))
- The contribution of one input quantity to the uncertainty of the result: ui(y) = |ci|·u(xi). When the sign matters (correlated inputs), the signed value is written ci·u(xi) (JCGM 100:2008, § 5.1.3).
- Input quantities with zero covariance, meaning they share no error source (JCGM 100:2008, § 5.1).
- Variance (u2)
- The square of a standard uncertainty. Variances of uncorrelated contributors can be added; standard uncertainties cannot (JCGM 100:2008, § C.3.2).
Note: JCGM 200:2012 is the International Vocabulary of Metrology (VIM). JCGM 100:2008 is the Guide to the Expression of Uncertainty in Measurement (GUM). NIST Technical Note 1297 gives NIST’s guidelines, which are consistent with the GUM.
References
Guides and Standards
- JCGM 100:2008, Evaluation of measurement data — Guide to the expression of uncertainty in measurement (GUM). Joint Committee for Guides in Metrology, 2008. https://www.bipm.org/documents/20126/2071204/JCGM_100_2008_E.pdf
- JCGM 101:2008, Evaluation of measurement data — Supplement 1 to the “Guide to the expression of uncertainty in measurement” — Propagation of distributions using a Monte Carlo method. Joint Committee for Guides in Metrology, 2008. https://www.bipm.org/documents/20126/2071204/JCGM_101_2008_E.pdf
- JCGM 200:2012, International vocabulary of metrology — Basic and general concepts and associated terms (VIM), 3rd edition. Joint Committee for Guides in Metrology, 2012. https://www.bipm.org/documents/20126/2071204/JCGM_200_2012.pdf
- BIPM, The International System of Units (SI), 9th edition (SI Brochure). Bureau International des Poids et Mesures, 2019. https://www.bipm.org/en/publications/si-brochure
- Thompson, A. and Taylor, B. N., Guide for the Use of the International System of Units (SI), NIST Special Publication 811, 2008 edition. National Institute of Standards and Technology. https://www.nist.gov/pml/special-publication-811
- Taylor, B. N. and Kuyatt, C. E., Guidelines for Evaluating and Expressing the Uncertainty of NIST Measurement Results, NIST Technical Note 1297, 1994 edition. National Institute of Standards and Technology. https://www.nist.gov/pml/nist-technical-note-1297
- Global ACI TECH-1-009, Global ACI Policy for Measurement Uncertainty in Calibration (formerly ILAC P14:09/2020, ILAC Policy for Measurement Uncertainty in Calibration). Global Accreditation Cooperation Incorporated. https://global-aci.org/en/global_aci-documents/?cat_id=5
- ISO/IEC 17025:2017, General requirements for the competence of testing and calibration laboratories. International Organization for Standardization. https://www.iso.org/standard/66912.html
- ISO 16063-21:2003, Methods for the calibration of vibration and shock transducers — Part 21: Vibration calibration by comparison to a reference transducer. International Organization for Standardization. https://www.iso.org/standard/27053.html
- ISO/TS 15530-4:2008, Geometrical Product Specifications (GPS) — Coordinate measuring machines (CMM): Technique for determining the uncertainty of measurement — Part 4: Evaluating task-specific measurement uncertainty using simulation. International Organization for Standardization. https://www.iso.org/standard/38696.html
- OIML R 111-1:2004, Weights of classes E1, E2, F1, F2, M1, M1–2, M2, M2–3 and M3 — Part 1: Metrological and technical requirements. International Organization of Legal Metrology. https://www.oiml.org/en/files/pdf_r/r111-1-e04.pdf
Books and Technical Notes
- Bell, S., A Beginner’s Guide to Uncertainty of Measurement, Measurement Good Practice Guide No. 11, Issue 2. National Physical Laboratory, 1999.
- Taylor, J. R., An Introduction to Error Analysis: The Study of Uncertainties in Physical Measurements, 2nd edition. University Science Books, 1997.
- Fluke Calibration, Technical Note 2011TN06B, Uncertainty Analysis for Flow Measured by molbloc-L and molbloc-S Mass Flow Transfer Standards. Fluke Calibration. https://media.fluke.com/c2a59e38-f9ba-409c-b1ba-b2e901015e1f_original%20file.pdf
Manufacturer Documentation
- Fluke Calibration, molbox1+ Specifications (A700K reference pressure monitor). https://www.fluke.com/en-us/product/calibration-tools/flow-calibration/gas-flow-standards/molbox1-flow-terminal
- Keysight Technologies, 3458A Digital Multimeter Data Sheet. https://www.keysight.com/us/en/assets/7018-06796/data-sheets/5965-4971.pdf
- Mitutoyo Corporation, Gauge Blocks Catalog (thermal expansion coefficient of steel gauge blocks). https://www.mitutoyo.com/webfoo/wp-content/uploads/E_section.pdf
Note: The symbol § is used in this article to represent the term section.
Originally published . Updated : rewritten to cover correlated input quantities, with step-by-step procedures, three worked examples, Excel formulas, common mistakes, and an expanded FAQ.